European Union

EU In Vitro Diagnostic Regulation (IVDR)

The framework for in vitro diagnostic devices in Europe, and the single largest regulatory shift the IVD sector has faced. Applicable since May 2022.

Instrument
Regulation (EU) 2017/746
Authority
Notified Bodies · European Commission · national competent authorities
The framework

What this market demands.

Regulation (EU) 2017/746 has applied since 26 May 2022. For the IVD sector the change was more severe than MDR was for medical devices: under the previous directive the large majority of IVDs could be self-certified, whereas under IVDR the great majority require notified body involvement.

The regulation replaced list-based classification with a rule-based system across four classes, A through D, determined by the risk the diagnostic result poses to the individual and to public health.

Performance evaluation under IVDR rests on three pillars: scientific validity, analytical performance and clinical performance. All three must be documented, maintained and updated across the product's life, which for many manufacturers is new work and not a reformatting of existing files.

Classification

Where a device sits decides what it costs.

Classification is the single most consequential determination in the process, and the one most often inherited as an assumption instead of being established as a conclusion.

Class A
Lowest risk: instruments, specimen receptacles, general laboratory products. Self-certified, unless supplied sterile, in which case a notified body is involved for the sterility aspects.
Class B
Moderate risk. Notified body involvement required, including assessment of technical documentation on a sampling basis.
Class C
High individual risk or moderate public health risk, including companion diagnostics and many infectious disease assays. Substantial notified body assessment.
Class D
Highest risk: agents posing a life-threatening risk with high propagation potential, and blood screening. Additional EU reference laboratory testing and expert panel scrutiny apply.
The pathway

The route to market, in order.

Steps can overlap, but they cannot meaningfully be reordered. Most delay comes from attempting one before the one before it is settled.

  1. Classify against the Annex VIII rules

    Determine the class from the intended purpose and the seven classification rules. Companion diagnostics and self-testing devices warrant particular care.

  2. Build the quality system

    A quality management system meeting Article 10, in practice built to ISO 13485 and subject to notified body audit for Classes B, C and D.

  3. Performance evaluation

    Assemble scientific validity, analytical performance and clinical performance evidence, documented in a performance evaluation report.

  4. Technical documentation

    Annex II and III documentation, including risk management, stability, and the performance evaluation plan and report.

  5. Conformity assessment

    Notified body assessment for Classes B, C and D. Class D additionally involves an EU reference laboratory and, where applicable, an expert panel.

  6. Post-market performance follow-up

    Operate PMS and post-market performance follow-up, with Periodic Safety Update Reports for Classes C and D.

Dates that matter

The calendar is part of the strategy.

Transitional deadlines have moved before and may move again. Treat these as orientation, not as a substitute for checking.

  1. 26 May 2022IVDR became applicable, replacing the In Vitro Diagnostic Directive.
  2. 31 December 2027End of the extended transition for Class D legacy devices, per Regulation (EU) 2024/1860.
  3. 31 December 2028End of the extended transition for Class C legacy devices.
  4. 31 December 2029End of the extended transition for Class B legacy devices and sterile Class A devices.
What goes wrong

The failures we see most often.

None of these are exotic. They are ordinary decisions that were reasonable at the time and expensive in hindsight.

  • Assuming self-certification carries over from the directive, which for most IVDs it does not
  • Treating performance evaluation as a literature summary instead of three distinct evidence pillars
  • Discovering companion diagnostic status late, having scoped the programme as Class B
  • Leaving notified body engagement until the technical file is finished, when capacity is the constraint
  • Relying on the extended transition dates without meeting the conditions attached to them

This page describes a public regulatory framework and was last reviewed 2026-08. It is provided for orientation and is not regulatory advice for a specific device. Requirements and deadlines change, so please confirm anything time-critical before relying on it.